Biological Safety Cabinet Class II A2: BSC-IIA2-1500. Lead Time: 4-6 Weeks. Features include high precision Air Velocity sensor, UV sterilization function, memory function in case of power-failure, audio and visual alarms, interlock functions, work area surrounded by negative pressure, and foot switch for front window height adjustment.
Lead Time: 4-6 Weeks
• Internal Size (W×D×H) mm: 1350×600×600mm
• External Size (W×D×H) mm: 1500x750x2250mm
• Tested Opening: Safety Height 200mm (8”)
• Max.Opening: 440mm
• Airflow System: 70% air recirculation, 30% air exhaust
• Downflow Velocity: 0.33±0.025m/s
• Inflow Velocity: 0.53±0.025m/s
• Total Air Exhaust: 500m3/h
• Filter Specification: Two, HEPA filter, ≥99.995% efficiency at 0.3 μm
• Supply HEPA Filter Size (mm): 1380x470x69
• Exhaust HEPA Filter Size (mm): 920x420x69
• Cleanliness Grade: Class 100 (ISO14644.1 Class 5)
• Display: Full color 7.0-inch touch screen
• Front Window: Motorized, Tempered Glass
• LED Lamp: 16W x2
• UV Lamp: 40W x1
• Rated Consumption: 1800W
• Illumination: Average Illuminance≥650 lx, the measured value of each illuminance≥ 430 lx
• Sound Emission: ≤67db
• Vibration Semi-peak Value: ≤5μm
• Power Supply: AC220V±20, 50Hz; 110V, 60HZ (Optional)
• Material: External Main Body: Cold rolling steel electrostatic spraying exterior, Work Surface: 304 stainless steel
• Net Weight: 320kg
• Gross Weight: 400kg
• Shipping Size(W×D×H) mm: 1630×880×2380
• Shipping Volume (m3): 3.5
Standard with high precision Air Velocity sensor. UV sterilization function and timing function can be set by appointment. With memory function in case of power-failure. Audio and visual alarm (Abnormal airflow velocity, HEPA filter failure, UV lamp failure, Front window at unsafe height). Interlock function: UV lamp and front window; UV lamp and blower, LED lamp; blower and front window. Work area surrounded by negative pressure, prevent leakage, improve Biosafety. Foot switch: adjust front window height, to avoid airflow turbulence caused by arm movement.
• High precision Air Velocity sensor
• UV sterilization function
• Memory function in case of power-failure
• Audio and visual alarms
• Interlock functions
• Work area surrounded by negative pressure
• Foot switch for front window height adjustment